Abstract:
The aim of this study was to test the measurement precision and accuracy of the Kajaani FS-100 giving attention to possible machine error in the measurements. Fiber length of pine pulps produced using polysulfide, kraft, biokraft and soda methods were determined using both FS-100 and FiberLab automated fiber length analyzers. The measured length values were compared for both methods. The measurement precision and accuracy was tested by replicated measurements using rayon stable fibers. Measurements performed on pulp samples showed typical length distributions for both analyzers. Results obtained from Kajaani FS-100 and FiberLab showed a significant correlation. The shorter length measurement with FiberLab was found to be mainly due to the instrument calibration. The measurement repeatability tested for Kajaani FS-100 indicated that the measurements are precise.