Species Identification of Clinical Coagulase-negative Staphylococci Isolated in Al-Shifa Hospital Gaza using Matrix-assisted Laser Desorption/Ionization-time of Flight Mass Spectrometry
Nahed Ali Al Laham

Abstract: Background: Staphylococcal species, notably, coagulase negative staphylococci (CoNS) are considered important causative agents of hospital-acquired infections (HAI) associated with catheters and implanted medical devices. The increasing incidence of CoNS in HAI confirms the need for an accurate and simple identification method at the species level. Here, the capacity of MALDI-TOF MS was compare in comparison of VITEK-2 and ID32STAPH to accurately identify clinical CoNS species and to determine its ability to correctly distinguish between methicillin resistant CoNS (MR-CoNS) and methicillin sensitive CoNS (MS-CoNS). Materials and Methods: In this study, MALDI-TOF MS, VITEK-2 and ID32STAPH was compared for phenotypic identification of CoNS at species level. Eighty one clinical isolates of CoNS representing six species that collected from different clinical samples were analyzed. All CoNS isolates were tested by catalase, Pastorex™, coagulase, nuc and mecA genes. Statistics were performed with Statistical Package for Social Sciences (SPSS®) program version 20 (Chicago, IL, USA). Results: Our results showed correct species identification by MALDI-TOF MS was obtained in 98.8% in comparison to 96.3 and 75.3% correct identification for the VITEK-2 and ID32STAPH, respectively. Only one strain was identified by MALDI-TOF MS at the genus level. In addition, MALDI-TOF MS identified staphylococcal to subspecies level including Staphylococcus hominis subsp., novobiosepticus and subsp., hominis, S. saprophyticus subsp., saprophyticus. Also the single strain of S. auricularis was only identified by MALDI-TOF MS. There were 60 (74.1%) MR-CoNS that showed mecA positive and 21 (25.9%) were MS-CoNS. The MALDI-TOF MS technique was unable to discriminate between MR-CoNS and MS-CoNS, because the topology of dendrogram generated from the spectra of MR-CoNS and MS-CoNS strains were almost the same. Conclusion: These results confirm the value of MALDI-TOF MS as simple, accurate and rapid method for phenotypic identification of clinical CoNS isolates. However, this technique was unable to discriminate between MR-CoNS and MS-CoNS strains.

Fulltext PDF Fulltext HTML

How to cite this article
Nahed Ali Al Laham , 2017. Species Identification of Clinical Coagulase-negative Staphylococci Isolated in Al-Shifa Hospital Gaza using Matrix-assisted Laser Desorption/Ionization-time of Flight Mass Spectrometry. Current Research in Bacteriology, 10: 1-8.

Keywords: Coagulase negative staphylococci, species identification, MALDI-TOF MS, VITEK-2 and ID32STAPH

INTRODUCTION

Hospital-acquired infections associated with catheters and implanted medical devices are most commonly caused by staphylococci, specially coagulase-negative staphylococci (CoNS). They are responsible for bacteremia, endocarditis, mediastinitis, meningitis and progressive joint destruction. Current methods for species identification are slow, costly and sometimes unreliable. To date there are 45 different Staphylococcus species with 24 subspecies. The clinical most relevant CoNS are Staphylococcus epidermidis, S. lugdunensis, S. saprophyticus and S. capitis1,2. The S. epidermidis is probably the major CoNS species causing nosocomial infections, although, other CoNS, such as S. lugdunensis, S. schleiferi and S. saprophyticus have been reported as even more pathogenic in a range of both nosocomial and community acquired infections3,4.

Identification of CoNS in the clinical laboratory can be problematic as many of the conventional methods and automated systems currently used do not always distinguish the different CoNS species reliably due to variable expression of biochemical characters. Moreover, molecular methods targeting the 16S rRNA, hsp60, femA, rpoB, gap, tuf and sodA are favored for diagnostic purposes but these methods often are expensive and time-consuming techniques2-6. Recently, matrix-assisted laser-desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) has shown to be a useful and simple method for the rapid identification of CoNS isolated in a routine clinical microbiology laboratory3-10. Starting from whole cells, cell lysates or bacterial extracts, it is possible to identify CoNS to species level within a few minutes based on mass analysis of the protein composition of the bacterial cell between 2,000 and 20,000 Da. A high quality reference database of spectra generated from well characterized strains and advanced analysis software for spectral pattern matching are essential to obtain accurate species identification. The output form the MALDI Biotyper, a frequently used MALDI-TOF MS instrument from Bruker® is a log score in the range of 0-3.0, computed by comparison of the peak list for an unknown isolate with the reference main spectra (MSP), containing information on average peak masses, average peak intensities and peak frequencies in the database. Bruker database version 3.1.1.0 includes 120 spectra of 37 different CoNS species11-15.

The aim of this study was to confirm the identification capacities of MALDI-TOF MS to accurately identify clinical CoNS species and to compare its capacity to the phenotypic VITEK-2 and ID32STAPH diagnostic methods. In addition, the study aimed to determine the ability of MALDI-TOF MS to distinguish between methicillin resistant CoNS (MR-CoNS) and methicillin sensitive CoNS (MS-CoNS).

MATERIALS AND METHODS

Bacterial isolates: For the present study, 81 unique non-duplicate CoNS isolates representing six species associated with diverse clinical infections that collected from different clinical samples including pus, urine, blood, wound, vaginal and ear swabs were collected from the largest public tertiary referral hospital (Al-Shifa Hospital) in Gaza. These isolates were collected in 2012, out of a total unique 1121 bacterial isolates (March 1-July 31). Isolates were obtained directly from the clinical laboratory of Al-Shifa Hospital and represent complete capture of all CoNS isolates during the stated collection period. Although, a defined sampling strategy was not employed, the collected strains likely reflect the clinical epidemiology of CoNS in Gaza, since Al-Shifa Hospital is the primary referral hospital for patients from all areas of the Gaza Strip. Well typed strains including nineteen reference strains were included (Table 1). Ethical approval from the Helsinki committee at the ministry of health in Gaza Strip (Approval No. PHRC/HC/36/15) was obtained for performing the current study.

Identification of CoNS and MR-CoNS: The CoNS isolates were identified phenotypically based on colonial morphology, Gram stain and by using the following methods: Catalase test, tube coagulase test, Pastorex™ Staph Plus latex agglutination (Bio-Rad, Hercules, California) and the Staph ID 32 API system (bioMérieux, France) according to the manufacturer’s instructions.

Table 1: Included reference strains

For VITEK-2 system, bacterial suspensions were prepared by suspending bacterial isolates in 0.45% saline to the equivalent of a 0.5-0.63 McFarland turbidity standard. The ID-GP identification card is a 64-well plastic card that includes 43 tests. Data were analyzed using VITEK-2 database, version 4.03 (bioMérieux, Marcy l’Etoile, France), according to the manufacturer’s instructions.

The identification of four CoNS isolates that gave discrepant results at the species level was obtained by sequencing an internal fragment of the sodA gene as previously described16,17. The partial sodA gene was amplified and the PCR product sequenced as previously described17. The nucleotide sequences were analyzed using the GenBank database of sodAint sequences of Staphylococcus type strains. A multiplex PCR assay was used for detection of the nuc gene to exclude S. aureus and mecA gene for detection of methicillin-resistance among CoNS isolates18.

MALDI-TOF MS: All isolates were streaked from stocks on tryptic soy agar (Lab M, Lancashire, UK) with 5% horse blood (E and O, Bonnybridge, Scotland) (TSA+B) and incubated overnight at 37°C. The MALDI-TOF MS analysis was performed on a pure subculture on TSA+B incubated overnight at 37°C in a CO2 incubator.

Only the direct transfer method was used. Briefly, one colony of each bacterial strain was spotted on a ground steel MALDI target plate. Each strain was spotted two times. Spots were allowed to dry at room temperature. Subsequently, 1 μL of matrix (Bruker Daltonik GmbH, Bremen, Germany), a saturated solution of α-cyano-4-hydroxycinnamic acid in 50% acetonitrile (Sigma-Aldrich) and 2.5% trifluoro-acetic acid (Sigma-Aldrich) was added to the spot, which was dried again. Afterwards, the target plate was introduced into the MALDI-TOF MS. Samples were analyzed with a Microflex LT MALDI-TOF MS instrument (Bruker Daltonik GmbH). Parameter settings were Ion Source (IS) 1 20 kV, IS2 18.5 kV, lens 8.5 kV, pulsed ion extraction 250 nucleotides and no gating. For spectral calibration, the Bruker® bacterial test standard (E. coli lysate) was used weekly. The measurements were performed in the automated mode using the software wizard (flex control version 3.0, MALDI Biotyper version 2.0 and MALDI Biotyper Automation Control version 2.0). Species identification was carried out with library version 3.1.1.0 containing MSP of 3290 reference strains (1805 different species). Ranking tables containing the 10 best matching MSP, sorted according to decreasing log score values, were obtained with the wizard. The acceptance criteria for identification results generally cited in the literature were used: A log score ≥1.7 and ≥2.0, respectively are the thresholds for a match at the genus and the species level. Scores of <1.7 were considered an unacceptable identification, according to the manufacturer’s recommendations. These thresholds were empirically determined4.

Statistics: Statistics were performed with Statistical Package for Social Sciences (SPSS®) program version 20 (Chicago, IL, USA). A p-value of less than 0.05 was considered significant.

RESULTS

Using the three phenotypic methods, all isolates were identified at the species level except four isolates. Discrepant results (n = 4) were reanalysed with MALDI-TOF MS. For confirmed discrepancies of these four CoNS, the sequencing of the sodA gene was performed and allowed the species identification of these four CoNS clinical isolates.

Species identification of reference strains: Reference strains (n = 19) were used to confirm the capacities of the MALD-TOF MS technology for species identification (Table 1). For all tested species at least one reference spectra was present in the Bruker database. MALDI-TOF MS was 100% correct in assigning the species. However, out of the 19 reference strains, six (S. auricularis, S. cohnii subsp., cohnii, S. pettenkoferi, S. saccharolyticus, S. schleiferi subsp., coagulans and S. schleiferi subsp., schleiferi) were identified only up to genus level and no reliable result was obtained for S. cohnii subsp., urealyticus.

Species identification of clinical CoNS strains: Eighty one clinical CoNS isolates were identified using MALDI-TOF MS, VITEK-2 system and ID32STAPH as described previously. Correct identification (ID) results were obtained for 98.8% (80/81), 96.3% (78/81) and 75.3% (61/81) with MALDI-TOF MS, VITEK-2 system and ID32STAPH, respectively. The correct ID for both MALDI-TOF MS and VITEK-2 system was significantly higher than the correct ID of ID32STAPH. Misidentification (Mis-ID) rates were 1.2% (1/81), 2.5% (2/81) and 18.5% (15/81), respectively. The percentage of strains with no-identification (no-ID) was 0.0% (0/81), 1.2% (1/81) and 6.2% (5/81), respectively. A correct ID (100%) of all strains in all three systems was not achieved. However with MALDI-TOF MS and VITEK-2 systems, it was obtained for three species, S. hemolyticus, S. epidermidis and S. warneri (Table 2).

Only MALDI-TOF MS identified staphylococcal to subspecies level including S. hominis subsp., novobiosepticus and subsp., hominis and S. saprophyticus subsp., saprophyticus. Also the single strain of S. auricularis was only correctly identified by MALDI-TOF MS and was confirmed by sequencing sodA gene PCR product. However, one of the S. hominis isolates could not be identified using MALDI-TOF MS.

For the CoNS species that most frequently cause diseases in humans or are isolated from human samples, S. hemolyticus (n = 33), S. epidermidis (n = 26), S. saprophyticus (n = 12) and S. hominis (n = 07), MALDI-TOF MS gave a correct identification in 98.7% (77/78), VITEK-2 system in 97.4% (76/78) and ID32STAPH in 78.2% (61/78) of these isolates (Table 2).

Differentiation of MR-CoNS from MS-CoNS strains using MALDI-TOF MS: According to the mecA PCR results there were 60 (74.1%) MR-CoNS and the rest (21 and 25.9%) were MS-CoNS. Figure 1 and 2 showed that, the topology of the dendrogram generated by the MALDI Biotyper 2.0 software from the spectra of MR-CoNS and MS-CoNS strains (with or without using formic acid as a matrix) were almost the same which make us unable to discriminate between them.

Table 2:
Identification obtained with the matrix-assisted laser desorption/ionization-time of flight mass spectrometry (MALDI-TOF MS), VITEK-2 system and ID32STAPH for the 81 clinical coagulase-negative staphylococci (CoNS) isolates
No.: Number of isolates, ID: Correct identification, Mis-ID: Misidentification, No-ID: No identification, *Four isolates of CoNS were identified at the species level by sequencing the sodA gene (reference standard), §Statistically significant difference in percentage of correct identification between the identification systems (p-value <0.05)

Fig. 1:
Dendrogram representation of hierarchical cluster analysis that generated by the MALDI Biotyper 2.0 software from the spectra of selected MR-CoNS and MS-CoNS strains (without using formic acid as a matrix)

Fig. 2:
Dendrogram representation of hierarchical cluster analysis that generated by the MALDI Biotyper 2.0 software from the spectra of selected MR-CoNS and MS-CoNS strains (with using formic acid as a matrix)

DISCUSSION

The CoNS are normal inhabitants of the human skin that lately have revealed themselves as significant etiological agents causing nosocomial infections, particularly in medical devices. The use of MALDI-TOF MS for staphylococcal identification is now considered routine in laboratories compared with the conventional phenotypical methods previously used. This technique allowed the identification of approximately 85% of the CoNS strains, whereas only 14% of the CoNS strains were identified to the species level with phenotypic methods4. Also, MALDI-TOF MS had previously demonstrated its powerful tool for the identification of clinically relevant species of CoNS3. Bacterial identification is routinely achieved using phenotypically based techniques. However, those techniques remain time-consuming and sometimes of limited value2-6, as for example for CoNS, where commercial identification kits identified only 37% of 177 CoNS isolates with the API 20 Staph system3.

In this study and as first aim, we compared between MALDI-TOF MS, VITEK-2 and ID32STAPH for the identification of CoNS to the species level. Our results showed correct species identification by MALDI-TOF MS was obtained in 98.8% compared to 96.3 and 75.3% correct identification for the VITEK-2 and ID32STAPH respectively. Dupont et al.2 achieved correct identification in 97.4, 79 and 78.6% using MALDI-TOF MS, phoenix and VITEK-2, respectively. In this study, Mis-ID rates were 1.2, 2.5 and 18.5%, respectively. The percentage of strains no-ID was 0.0, 1.2 and 6.2%, respectively. Comparable findings were achieved in Dupont et al.2 study where they found that Mis-ID occurred in 1.3, 21 and 10.3% with MALDI-TOF MS, Phoenix and VITEK-2, respectively and the rates of no-ID results were low with all systems: 1.3, 0 and 0.9% with MALDI-TOF-MS, phoenix and VITEK-2, respectively. However, Delmas et al.19 obtained a correct identification of only 71.2% for clinical and environmental CoNS with the VITEK-2 ID system and 82.5% when isolates belonging to species absent from the automate database were excluded. Nevertheless, overall identification accuracy for CoNS with VITEK-2 was in the range of 91-9620-22. In another study using MALDI-TOF MS, a total of 151 strains out of 152 (99.3%) CoNS were correctly identified at the species level and only one strain was identified at the genus level6. Spanu et al.23 evaluated the use of MALDI-TOF MS for the identification of 450 blood isolates of the most relevant staphylococcal species, using sequence analysis of the rpoB gene as the reference method. A correct species identification by MALDI-TOF MS was obtained in 99.3%, with only three isolates (0.7%) being misidentified. These findings are comparable to our results (correct species identification: 98.8%, misidentification: 1.2%).

In this study, MALDI-TOF MS correctly identified all six CoNS species to the species level, except for S. hominis (85.7%). Many studies showed comparable results, however, Dupont et al.2 showed same results except for S. hemolyticus (85.8%). Only MALDI-TOF MS identified staphylococcal to subspecies level including S. hominis subsp., novobiosepticus and subsp., hominis and S. saprophyticus subsp., saprophyticus. Same results were obtained by Spanu et al.23 where MALDI-TOF MS correctly identified all the staphylococcal subspecies studied, including S. capitis subsp., capitis and subsp., urealyticus, S. cohnii subsp., urealyticus, S. hominis subsp., novobiosepticus and subsp., hominis, S. saprophyticus subsp., saprophyticus, S. schleiferi subsp., schleiferi and S. sciuri subsp., sciuri23.

The second aim of this report was to determine the ability of MALDI-TOF MS to correctly distinguish between MR-CoNS and MS-CoNS. The topology of dendrogram generated from the spectra of MR-CoNS and MS-CoNS strains were almost the same which make us unable to discriminate between them. The molecular available methods used to type strains, including pulsed-field gel electrophoresis and multilocus sequence typing are laborious and expensive to perform. Several studies have addressed the use of MALDI-TOF MS as a method for typing methicillin-resistant S. aureus24-26, but there is no any previous report studying CoNS. While some studies showed the ability of MALDI-TOF MS to distinguish between methicillin-susceptible and resistant S. aureus27,28, contradictory reports have been published on this issue24,29. Yet, the MALDI-TOF MS method revealed different clonal lineages of S. epidermidis that were of either human or environmental origin, which suggests that the MALDI-TOF MS method could be useful in the profiling of staphylococcal strains6. On the other hand, Bittar et al.30 showed that MALDI-TOF MS can distinguish between positive and negative PVL and TSST S. aureus isolates by producing specific peaks of 4448 and 5302 m/z. But Szabados et al.31 indicate that the peaks of 4448 and 5302 m/z are independent of the presence of PVL and TSST as they were found in all 104 MRSA strains with different PFGE in contrast to the aforementioned report. Moreover, they did not find evidence for an association of single protein peaks in the m/z range of 3000-10,000 Da with the presence of the toxins sea, seb, sec, sed, see, seg, seh, sei and sej. So, they suggest that a MALDI-TOF MS-based typing approach in PVL-positive S. aureus is clearly inferior and inapplicable compared to well-established molecular typing methods31.

Finally and because MALDI-TOF MS is easy to perform, fast and relatively cheap, it is the preferred method for the species identification of clinical CoNS in routine clinical microbiology5,11,12.

CONCLUSION

Many previously reported studies had demonstrated that MALDI-TOF MS is a powerful tool for the identification of clinically relevant species of CoNS and showed that it is a simple, accurate and rapid method for phenotypic identification of clinical CoNS isolates. Our results obtained in this study demonstrate the higher performance of MALDI-TOF MS in the identification of CoNS to the species level and even more to the subspecies level compared to VITEK-2 and ID32STAPH. However, this study revealed for the first time the inability of MALDI-TOF MS to correctly distinguish between MR-CoNS and MS-CoNS.

ACKNOWLEDGMENTS

The researchers would like to thank the German Academic Exchange Service (DAAD), Germany and the Erasmus Mundus External Cooperation Window (EM ECW lot III), Belgium, EU for support of my work during my visits to the University Clinic of Muenster, Germany and the Free University of Brussels, Belgium respectively. Also my thanks to the laboratory technologists at Al Shifa hospital for their kind help.

REFERENCES

  • Murray, P.R., 2010. Matrix‐assisted laser desorption ionization time‐of‐flight mass spectrometry: Usefulness for taxonomy and epidemiology. Clin. Microbiol. Infect., 16: 1626-1630.
    CrossRef    Direct Link    


  • Dupont, C., V. Sivadon‐Tardy, E. Bille, B. Dauphin and J.L. Beretti et al., 2010. Identification of clinical coagulase‐negative staphylococci, isolated in microbiology laboratories, by matrix‐assisted laser desorption/ionization‐time of flight mass spectrometry and two automated systems. Clin. Microbiol. Infect., 16: 998-1004.
    CrossRef    Direct Link    


  • Carbonnelle, E., J.L. Beretti, S. Cottyn, G. Quesne, P. Berche, X. Nassif and A. Ferroni, 2007. Rapid identification of Staphylococci isolated in clinical microbiology laboratories by matrix-assisted laser desorption ionization-time of flight mass spectrometry. J. Clin. Microbiol., 45: 2156-2161.
    CrossRef    Direct Link    


  • Argemi, X., P. Riegel, T. Lavigne, N. Lefebvre and N. Grandpre et al., 2015. Implementation of matrix-assisted laser desorption ionization-time of flight mass spectrometry in routine clinical laboratories improves identification of coagulase-negative staphylococci and reveals the pathogenic role of Staphylococcus lugdunensis. J. Clin. Microbiol., 53: 2030-2036.
    CrossRef    Direct Link    


  • Carpaij, N., R.J. Willems, M.J.M. Bonten and A.C. Fluit, 2011. Comparison of the identification of coagulase-negative staphylococci by matrix-assisted laser desorption ionization time-of-flight mass spectrometry and tuf sequencing. Eur. J. Clin. Microbiol. Infect. Dis., 30: 1169-1172.
    CrossRef    Direct Link    


  • Dubois, D., D. Leyssene, J.P. Chacornac, M. Kostrzewa and P.O. Schmit et al., 2010. Identification of a variety of Staphylococcus species by matrix-assisted laser desorption ionization-time of flight mass spectrometry. J. Clin. Microbiol., 48: 941-945.
    CrossRef    Direct Link    


  • Clark, AE., E.J. Kaleta, A. Arora and D.M. Wolk, 2013. Matrix-assisted laser desorption ionization-time of flight mass spectrometry: A fundamental shift in the routine practice of clinical microbiology. Clin. Microbiol. Rev., 26: 547-603.
    CrossRef    Direct Link    


  • Bizzini, A. and G. Greub, 2010. Matrix‐assisted laser desorption ionization time‐of‐flight mass spectrometry, a revolution in clinical microbial identification. Clin. Microbiol. Infect., 16: 1614-1619.
    CrossRef    Direct Link    


  • Bizzini, A., C. Durussel, J. Bille, G. Greub and G. Prod'hom, 2010. Performance of matrix-assisted laser desorption ionization-time of flight mass spectrometry for identification of bacterial strains routinely isolated in a clinical microbiology laboratory. J. Clin. Microbiol., 48: 1549-1554.
    CrossRef    Direct Link    


  • Weinstein, M.P., L.B. Reller and R. Patel, 2013. Matrix-assisted laser desorption ionization-time of flight mass spectrometry in clinical microbiology. Clin. Infect. Dis., 57: 564-572.
    CrossRef    Direct Link    


  • Han, H.W., H.C. Chang, A.H. Hunag and T.C. Chang, 2015. Optimization of the score cutoff value for routine identification of Staphylococcus species by matrix-assisted laser desorption ionization-time-of-flight mass spectrometry. Diagn. Microbiol. Infect. Dis., 83: 349-354.
    CrossRef    Direct Link    


  • Lee, M., H.S. Chung, H.W. Moon, S.H. Lee and K. Lee, 2015. Comparative evaluation of two matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) systems, Vitek MS and Microflex LT, for the identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories. J. Microbiol. Methods, 113: 13-15.
    CrossRef    Direct Link    


  • Martiny, D., L. Busson, I. Wybo, R.A. El Haj, A. Dediste and O. Vandenberg, , 2012. Comparison of the microflex LT and vitek MS systems for routine identification of bacteria by matrix-assisted laser desorption ionization-time of flight mass spectrometry. J. Clin. Microbiol., 50: 1313-1325.
    CrossRef    Direct Link    


  • De Carolis, E., A. Vella, L. Vaccaro, R. Torelli and T. Spanu et al., 2014. Application of MALDI-TOF mass spectrometry in clinical diagnostic microbiology. J. Infect. Dev. Countries, 8: 1081-1088.
    CrossRef    Direct Link    


  • Dingle, T.C. and S.M. Butler-Wu, 2013. MALDI-TOF mass spectrometry for microorganism identification. Clin. Lab. Med., 33: 589-609.
    CrossRef    Direct Link    


  • Poyart, C., G. Quesne, C. Boumaila and P. Trieu-Cuot, 2001. Rapid and accurate species-level identification of coagulase-negative staphylococci by using the sodA gene as a target. J. Clin. Microbiol., 39: 4296-4301.
    CrossRef    Direct Link    


  • Sivadon, V., M. Rottman, J.C. Quincampoix, V. Avettand and S. Chaverot et al., 2004. Use of sodA sequencing for the identification of clinical isolates of coagulase‐negative staphylococci. Clin. Microbiol. Infect., 10: 939-942.
    CrossRef    Direct Link    


  • Becker, K., R. Rotn and G. Peters, 1998. Rapid and specific detection of toxigenic Staphylococcus aureus: Use of two multiplex PCR enzyme immunoassays for amplification and hybridization of staphylococcal enterotoxin genes, exfoliative toxin genes and toxic shock syndrome toxin 1 gene. J. Clin. Microbiol., 36: 2548-2553.
    Direct Link    


  • Delmas, J., J.P. Chacornac, F. Robin, P. Giammarinaro, R. Talon and R. Bonnet, 2008. Evaluation of the Vitek 2 system with a variety of Staphylococcus species. J. Clin. Microbiol., 46: 311-313.
    CrossRef    Direct Link    


  • Spanu, T., M. Sanguinetti, D. Ciccaglione, T. DInzeo, L. Romano, F. Leone and G. Fadda, 2003. Use of the VITEK 2 system for rapid identification of clinical isolates of Staphylococci from bloodstream infections. J. Clin. Microbiol., 41: 4259-4263.
    CrossRef    Direct Link    


  • Funke, G. and P. Funke-Kissling, 2005. Performance of the new VITEK 2 GP card for identification of medically relevant gram-positive cocci in a routine clinical laboratory. J. Clin. Microbiol., 43: 84-88.
    CrossRef    Direct Link    


  • Ligozzi, M., C. Bernini, M.G. Bonora, M. de Fatima, J. Zuliani and R. Fontana, 2002. Evaluation of the VITEK 2 system for identification and antimicrobial susceptibility testing of medically relevant gram-positive cocci. J. Clin. Microbiol., 40: 1681-1686.
    CrossRef    Direct Link    


  • Spanu, T., E. de Carolis, B. Fiori, M. Sanguinetti, T. D'Inzeo, G. Fadda and B. Posteraro, 2011. Evaluation of matrix-assisted laser desorption ionization-time-of-flight mass spectrometry in comparison to rpoB gene sequencing for species identification of bloodstream infection staphylococcal isolates. Clin. Microbiol. Infect., 17: 44-49.
    CrossRef    Direct Link    


  • Wolters, M., H. Rohde, T. Maier, C. Belmar-Campos and G. Franke et al., 2011. MALDI-TOF MS fingerprinting allows for discrimination of major methicillin-resistant Staphylococcus aureus lineages. Int. J. Med. Microbiol., 301: 64-68.
    CrossRef    Direct Link    


  • Jackson, K.A., V. Edwards-Jones, C.W. Sutton and A.J. Fox, 2005. Optimisation of intact cell MALDI method for fingerprinting of methicillin-resistant Staphylococcus aureus. J. Microbiol. Methods, 62: 273-284.
    CrossRef    Direct Link    


  • Walker, J., A.J. Fox, V. Edwards-Jones and D.B. Gordon, 2002. Intact cell mass spectrometry (ICMS) used to type methicillin-resistant Staphylococcus aureus: Media effects and inter-laboratory reproducibility. J. Microbiol. Methods, 48: 117-126.
    CrossRef    Direct Link    


  • Edwards-Jones, V., M.A. Claydon, D.J. Evason, J. Walker, A.J. Fox and D.B. Gordon, 2000. Rapid discrimination between methicillin-sensitive and methicillin-resistant Staphylococcus aureus by intact cell mass spectrometry. J. Med. Microbiol., 49: 295-300.
    CrossRef    Direct Link    


  • Majcherczyk, P.A., T. McKenna, P. Moreillon and P. Vaudaux, 2006. The discriminatory power of MALDI-TOF mass spectrometry to differentiate between isogenic teicoplanin-susceptible and teicoplanin-resistant strains of methicillin-resistant Staphylococcus aureus. FEMS Microbiol. Lett., 255: 233-239.
    CrossRef    Direct Link    


  • Bernardo, K., N. Pakulat, M. Macht, O. Krut and H. Seifert et al., 2002. Identification and discrimination of Staphylococcus aureus strains using matrix‐assisted laser desorption/ionization‐time of flight mass spectrometry. Proteomics, 2: 747-753.
    CrossRef    Direct Link    


  • Bittar, F., Z. Ouchenane, F. Smati, D. Raoult and J.M. Rolain, 2009. MALDI-TOF-MS for rapid detection of staphylococcal Panton-Valentine leukocidin. Int. J. Antimicrob. Agents, 34: 467-470.
    CrossRef    Direct Link    


  • Szabados, F., K. Becker, C. von Eiff, M. Kaase and S. Gatermann, 2011. The matrix-assisted laser desorption/ionisation time-of-flight mass spectrometry (MALDI-TOF MS)-based protein peaks of 4448 and 5302 Da are not associated with the presence of Panton-Valentine leukocidin. Int. J. Med. Microbiol., 301: 58-63.
    CrossRef    PubMed    Direct Link    

  • © Science Alert. All Rights Reserved