INTRODUCTION
Water quality is assessed in terms of the amount of organic compounds present
in it. Biochemical Oxygen Demand (BOD) is an important and widely used parameter
for water bodies pollution monitoring (Akpoveta et al.,
2011; Alquwaizany et al., 2011; Aluyi
et al., 2006; Ayeni et al., 2011;
Khan et al., 2003; Mahre et
al., 2007; Norizan et al., 2011; Ogunfowokan
et al., 2005; Zainudin et al., 2010).
The BOD load of wastewater needs to be determined to ensure it is within the
parameter limits of effluents that can be discharged into inland waters under
the Sewage and Industrial Effluents Regulations 1979 (How,
2003). According to the American Public Health Association (APHA,
1992), the standard method for BOD5 assay requires 5 days of
incubation time at ±20°C in the dark. This conventional method of
BOD measurement is time consuming, irreproducible and labour intensive with
questionable accuracy. Measurement depends on temperature, oxygen concentration,
presence of toxins, as well as the type, quantity and quality of seeding microorganism
(Pasco et al., 2000, 2004).
The long duration of BOD5 assay is not suitable for on-line monitoring
especially where rapid feedback is essential for environmental monitoring and/or
process control (Morris, 2005).
To overcome the shortcomings in the traditional BOD5, various rapid
microbial biosensors have been developed. Microbial BOD biosensor was first
discovered by Karube et al. (1977), where the
sensor consists of a combination of an oxygen electrode as a transducer and
a bio-film (Nakamura et al., 2008). Microorganisms
immobilized in the bio-film were placed in close, intimate contact with an amperometric
oxygen electrode. When organic compounds present in the samples are degraded,
dissolved oxygen is consumed. Dissolved oxygen consumed is proportional to the
amount of biodegradable organics and designated as BOD5 value.
To date, most BOD biosensors still focus on measuring oxygen uptake by immobilized
cells attached to an oxygen electrode (Riedel et al.,
1990; Tan and Wu, 1999; Lehmann
et al., 1999; Liu et al., 2000). The
history of BOD sensor developments has been well summarized in two literatures
(Nakamura et al., 2007, 2008).
But these BOD sensors are limited either by the availability of oxygen, or by
pure microbial cultures with a narrow substrate range, or they require calibration
to a BOD5 standard solution.
The conventional 5-day BOD assay can achieve an average of 60.5% of organic
compounds degradation while using Glucose-Glutamic Acid (GGA) as the standard
check solution (Pasco et al., 2000; Morris
et al., 2001). For the classic microbial BOD sensor, only less than
1% of the organic compounds are degraded (Morris et al.,
2001). To increase the percentage degradation of organic compounds for rapid
BOD biosensor, a novel technique employing mediator (redox dye) such as ferricyanide
ion and an amperometric transducer was introduced by Pasco
et al. (2000). This technique replaces oxygen as final electron acceptor
for the biochemical reaction in the detection process. More recently, this kind
of mediated microbial sensors have received much attention for rapid BOD measurement
(Yoshida et al., 2000, 2001;
Catterall et al., 2001, 2003;
Trosok et al., 2001; Morris
et al., 2001; Pasco et al., 2004;
Morris, 2005).
Ferricyanide is an efficient redox mediator to shuttle electrons from the redox center of reduced microbial enzymes to the electrode in the presence of excess Glucose/Glutamic Acid (GGA). Thus, ferricyanide can be used instead of oxygen as an electron acceptor during microbial catabolism of organic compounds. The microbially reduced ferricyanide is then quantified electrochemically upon re-oxidation at the electrode surface. The amount of ferricyanide reduced is directly proportional to the amount of organic material present in a sample. The mechanism of this process is represented in the following equation:
Ferricyanide is about 10,000 times more soluble than oxygen in water. Moreover,
higher population of microbial cells can be used without the depletion of the
electron acceptor. Hence, large microbial consumption of organic matter can
be achieved in much shorter time (Pasco et al., 2000;
Morris et al., 2001).
FM-BOD (Ferricyanide-mediated determination of BOD) approaches that have been
reported can be divided into two general types. The first type is the bioreactor
assay that includes a separate incubation and detection step (Pasco
et al., 2000). The second type is ferricyanide-mediated amperometric
BOD biosensors in which the microbial was immobilized directly on the tip of
the working electrode (Yoshida et al., 2000,
2001; Trosok et al., 2001).
The first approach was employed by Pasco et al.
(2000), where it is named as MICREDOX®. In this study, the
first approach was employed too and it is named as FM-BOD assay. FM-BOD assay
comprises of two steps that occur independently. Firstly, FM-BOD assay involves
incubation of the sample solution which consists of microorganisms, ferricyanide
and the sample organic substrate. In the second step, detection of accumulation
of the reduced form of mediator (ferrocyanide) or the decrease in the oxidized
form of mediator (ferricyanide) is involved. Pasco et
al. (2000) quantified the microbially reduced mediator by measuring
the charge required for its re-oxidation with a coulometric transducer. Coulometric
detection was found to be labour-intensive and time consuming. Thus, several
improvements of BOD detection have been reported afterwards Morris
et al. (2001), Catterall et al. (2001,
2003), Pasco et al. (2004)
and Morris (2005).
The second method is the ferricyanide-mediated amperometric BOD biosensors
described by Yoshida et al. (2000, 2001)
and Trosok et al. (2001). This FM-BOD biosensor
method is similar with traditional BOD biosensors in which the biocatalyst is
immobilized on the tip of the working electrode to eliminate the entry of biocatalyst
into the sample solution. The difference is that the detection principle involves
microbially produced ferrocyanide being re-oxidized back to ferricyanide at
the electrode surface. The FM-BOD biosensor needs to be calibrated to a standard
organic solution before analyzing.
Though offering simple and rapid monitoring of BOD, FM-BOD biosensor has its limitations. Firstly, sensor fouling may occur due to constant contact of samples with the sensor resulting in constant sensor recalibration. The total BOD value may also be overestimated since only the most easily biodegradable portion of a sample will be degraded in the short reaction time. Furthermore, due to the variations in biodegradable organics content between samples frequent recalibration of the BOD biosensor is needed.
To demonstrate the feasibility of the FM-BOD approach, this study sought to employ a locally developed microbial consortium as biocatalyst to measure the BOD load of synthetically prepared organic solutions and real industrial waste water. The use of acclimated microorganisms is expected to extend the application of the FM-BOD assay to a broad range of wastewater.
MATERIALS AND METHODS
Reagents: BOD standard solution, Glucose Glutamic Acid (GGA) (150 mg
L-1 glucose, 150 mg L-1 glutamic acid) was prepared according
to APHA procedures (APHA, 1992). This solution was assigned
a BOD value of 198±30 mg BOD5-1. Potassium ferricyanide
solutions (analytical reagent grade) were prepared in Phosphate Buffer Saline
(PBS), pH 7. The Organization for Economic Cooperation and Development (OECD)
synthetic wastewater solution comprised of 16 g L-1 peptone, 11 g
L-1 meat extract and 3 g L-1 urea (diluted to100-fold)
(Morris, 2005). All solutions were prepared in deionized
water.
Microbial consortium preparation: The microbial consortium used in this study comprised of microorganisms isolated from various environmental sources in Malaysia. Microorganisms were grown in Tryptic Soy Broth (TSB) at 37°C in a shaking incubator for 16-18 h. Cells were then harvested by centrifugation at 4000 rpm for 15 min at room temperature, washed twice with phosphate buffer (pH 7) and then resuspended in phosphate buffer (pH 7). Cells were adjusted to a final absorbance of 5 at 600 nm using a 100 VIS Spectrophotometer (BUCK Scientific).
Sample preparation: Three sample solutions: blank, endogenous control and real sample were prepared in a final volume f 10 mL in serum bottles. Real sample contained 5 mL of appropriate bacterial suspension (OD600 5), 1.5 mL of 400 mM potassium ferricyanide and 3.5 mL of standard GGA solution/OECD solution (or real wastewater). Blanks were prepared by replacing ferricyanide with Phosphate Buffer Saline (PBS). Endogenous control solutions were prepared by replacing GGA standard solution with sterile deionized water. All solutions were spurged with oxygen free nitrogen for 15 min. For real samples, two further controls were employed: (1) Sample in the absence of ferricyanide and microorganisms to determine if there were any species present that undergo electrochemical oxidation and (2) Sample in the absence of microorganisms to determine if there were any species present capable of reducing ferricyanide to ferrocyanide.
Incubation: The samples prepared were incubated at 37°C in a shaking water bath. Sample (1 mL) was removed at hourly intervals for 2 h and the microbial reaction was terminated by centrifugation at 14,000 rpm for 15 min. The supernatant solution was then added into Phosphate Buffered Saline (PBS) to a total volume of 30 mL and bubbled with nitrogen gas for 30 sec before being analysed for microbially produced ferrocyanide using voltammetry.
Voltammetric detection: Measurements were conducted using a linear sweep voltammetry at potential 0-600 mV versus Ag/AgCl reference electrode. A 10 μm platinum microelectrode was used as the working electrode and platinum gauze auxiliary electrode was employed to complete the three electrode electrochemical cell. The limiting current value obtained for a given sample at a given incubation time was taken to determine the ferrocyanide concentration and thus the amount of organics degraded.
Calculation of FM-BOD5 equivalent values: As in BOD5
assay, GGA solution was used as a standard check solution for the FM-BOD assay.
In all cases, the limiting current values of the endogenous control solution
were subtracted from the sample and GGA limiting current values prior to calculation.
Limiting current (ilim) values obtained throughout the FM-BOD incubation
were divided by values obtained from GGA standard solution. This value is termed
the normalized limiting current and is a dimensionless parameter. This parameter
can be converted to a FM-BOD5 equivalent value by multiplication
of 198 mg L-1 (average accepted BOD5 value for the GGA
solution) (Catterall et al., 2003; Morris,
2005).
When the OECD synthetic wastewater standard was used as calibration standard,
limiting current (ilim) values obtained throughout the FM-BOD incubation
were divided by values obtained from OECD standard solution. In all cases, the
limiting current values of the endogenous control solution were subtracted from
the sample and OECD limiting current values prior to calculation. This normalized
limiting current was converted to a FM-BOD5 equivalent value by multiplication
of 170 mg L-1 (the average accepted BOD5 value for the
OECD solution following 100-fold dilution) (Morris, 2005).
Statistical analysis: Experiments were typically done in triplicates and the standard deviation of the mean values were calculated accordingly. The analysis of data were conducted at 5% level of confidence (p<0.05).
RESULTS AND DISCUSSION
Use of Ferricyanide by microbial consortium and commercial consortia for
GGA degradation: Locally developed microbial consortium and commercial consortia
were tested for their ability to use ferricyanide for degradation of GGA. Apart
from its use in standard BOD5 assay, GGA is also the most commonly
employed calibration standard for BOD biosensors. The locally developed microbial
consortium mainly consists of four bacteria species from various environmental
sources in Malaysia. The advantage of the microbial consortium is that it has
bacteria from various sources, thus could degrade greater type of organic compounds
in the real wastewater samples (Liu et al., 2000;
Liu and Mattiasson, 2002).
A commercial consortium was assessed to compare the response of our locally developed microbial consortium in the FM-BOD assay. The limiting current values obtained from repeated application of commercial consortium in FM-BOD assay were very low and not reproducible compared to our locally developed microbial consortium. This is because commercial consortium is already premixed and it is grown in TSB and adjusted to OD600 5, the exact concentration of each bacterium is unknown. On the contrary, our locally developed microbial consortium was adjusted to OD600 5 prior to mixing. Each bacterium has the same concentration and reproducibility of results was proven in repeated test. The results indicate that commercial consortium was not able to utilize ferricyanide as the final electron acceptor in the biochemical reaction for organic biodegradation.
From Fig. 1, a commercial consortium was replicated in this
study to compare with locally developed microbial consortium. It can be seen
that the limiting current response obtained for the microbial consortium is
higher compared to both commercial consortiums at the first and second hour
incubation. This indicates that the microbial consortium has rapidly degraded
GGA in the first and second hour. It is also shown that significant degradation
of GGA can already be observed in the first hour, showing that a rapid BOD detection
within 1 h is possible.
|
Fig. 1: |
Limiting current values obtained at various incubation times
for locally developed microbial consortium and commercial consortium in
GGA standard solution. Each point represents the average of response from
three replicate measurements. Endogenous control values have been subtracted.
Microbial consortium final absorbance = 2.5. Ferricyanide final concentration
= 60 mM. Limiting currents determined by voltammetry at Eapp
= +450 mV (vs Ag/AgCl) |
From the limiting current values, the extent of GGA degradation can be determined.
The extent of GGA degradation, expressed as percent conversion, was calculated
as previously described by Morris (2005). The data of
microbial consortium in Fig. 1 indicates that more than 45%
of the GGA solution had been degraded in the first hour of incubation. This
compares favourably with both the BOD5 assay (~60% GGA degraded in
5 days) and that previously reported by Morris et al.
(2005) (~40% GGA degraded in 1 h). Hence, in the FM-BOD method, the BOD
of a sample can be determined in 1 h compared to 5 days with the standard BOD5
assay.
Linear dynamic range for standard calibration solutions: GGA standard
solution and OECD synthetic solution were tested for their ability as a calibration
solution for the FM-BOD assay in 1 h incubation. GGA is a standard solution
for the conventional BOD5 assay and also used extensively for BOD
biosensors. However, OECD synthetic solution is the preferred calibration solution
for BOD biosensors due to more complex and low degradable organic compounds
(Liu et al., 2000). OECD synthetic solution is
made synthetically to mimic the real sample.
Figure 2 shows that the limiting current values are directly
proportional to the substrate concentration at approximately 400 mg L-1
for GGA solution and to approximately 250 mg L-1 for the OECD solution.
Comparing both lines, OECD shows a lower limiting current value which is indicative
of slower biodegradation rate. This explains that although OECD synthetic wastewater
is often the preferred calibration solution, most commonly employed calibration
solution for BOD biosensors is GGA (Morris, 2005). In
addition, in rapid BOD biosensor, only certain amount of assimilable organic
compounds in real samples can be degraded in such a short period. GGA with only
two simple components meets this requirement. Consequently, GGA with better
performance was chosen as the calibration solution for the experiments that
follow. Real wastewater samples were diluted to approximately 200 mg L-1
prior to starting FM-BOD assay.
|
Fig. 2: |
Limiting current values obtained at various substrate concentrations
for mixed microbial consortium in GGA standard solution and OECD synthetic
wastewater. Each point represents the average of response from three replicate
measurements. Endogenous control values have been subtracted. Microbial
consortium final absorbance = 2.5. Ferricyanide final concentration = 60
mM. Limiting currents determined by voltammetry at Eapp = +450
mV (vs Ag/AgCl) |
|
Fig. 3: |
Limiting current values obtained at various microbial concentrations
for mixed microbial consortium in GGA standard solution. Each point represents
the average response of three replicate measurements. Endogenous control
values have been subtracted. Microbial consortium final absorbance = 0.5,
1.5, 2.5, 3.5, 4.5. Ferricyanide final concentration = 60 mM. Limiting currents
determined by voltammetry at Eapp = +450 mV (vs Ag/AgCl) |
Effect of concentration of the mixed microbial consortium: The influence of microbial concentration towards the BOD measurements is important as it could affect the biochemical reaction. For instance, when microbial population is too low, only a small amount of organic substrate in the samples could be oxidized, thus making the biocatalyst a limiting factor. On the contrary, when the microbial population is too high, electron acceptor (oxygen in BOD5, ferricyanide for FM-BOD) may become the rate limiting reactant in order to fully degrade the organic substrate. Therefore, microbial with appropriate concentration is vital for the BOD measurement in this study.
Figure 3 shows the limiting current values obtained at various
concentration of mixed microbial consortium after incubation of one hour with
the standard GGA solution in the presence of ferricyanide. From the plot, it
is observed that the limiting current values increased gradually at concentrations
lower than OD600 2.5, before reaching a stable limiting current of
approximately 10 nA at concentration above 2.5. The optimum concentration with
a final absorbance of OD600 2.5 was then selected for the remaining
study and this finding is in agreement with previous study by Morris
(2005).
Effect of mediator concentration: As previously described, solubility of oxygen in water is low and is a limiting reactant for BOD5. Oxygen which acts as final electron acceptor is rapidly reaction rate, ferricyanide which is approximately 10 000 times more soluble than oxygen was employed in FM-BOD assay. This approach is used to shorten the reaction time for the purpose of rapid BOD detection. The effect of varying the ferricyanide concentration was investigated after incubation of one hour in GGA standard solution.
As shown in Fig. 4, limiting current values were dependent
on ferricyanide concentration below 40 mM and independent above that. The reaction
of ferricyanide concentration below 40 mM was limited by the availability of
electron acceptor. Above 40 mM, the reaction was limited by substrate concentration
which higher concentration of substrate is needed to reduce ferricyanide microbiall
(Morris et al., 2001). Ferricyanide concentration
with 60 mM was employed in all subsequent experiments to ensure that ferricyanide
does not limit the microbial process.
|
Fig. 4: |
Limiting current values obtained at various ferricyanide
concentrations for mixed microbial consortium in GGA standard solution.
Each point represents the average of response from three replicate measurements.
Endogenous control values have been subtracted. Microbial consortium final
absorbance = 2.5. Ferricyanide final concentration = (20, 40, 60, 80, 10)
mM. Limiting currents determined by voltammetry at Eapp = +450
mV (vs Ag/AgCl) |
Table 1: |
Ferricyanide-mediated BOD5 equivalent values for organic
standard solutions |
 |
Each point represents the average of response from three replicate
measurements. Endogenous control values have been subtracted and the BOD5
equivalent values have been calculated by comparison with the GGA standard.
Microbial consortium final absorbance = 2.5. Ferricyanide final concentration
= 60 mM. Limiting currents determined by voltammetry at Eapp
= +450 mV (vs Ag/AgCl). *Catterall et al. (2001)
and Morris et al. (2001) |
Application of FM-BOD assay to organic standard solution: The purpose
of this study is to determine the ability of the microbial consortium together
with FM-BOD assay in predicting BOD5 values of several organic solutions
(glucose, sucrose, glutamic acid and glycine). All organic standard solutions
were prepared at concentrations equivalent to BOD5 values of 200mg
BOD5-1. Limiting current values were obtained voltametrically
and converted to FM-BOD5 equivalent values by assigning general standard
solution BOD198 GGA as calibration standard. Table
1 shows FM-BOD5 equivalent values obtained from microbial consortium
employed in various organic solution after 1, 2 and 3 hour incubation. The FM-BOD5
equivalent values of microbial consortium for the simple sugars glucose and
sucrose were within the ±15% of the standard 5-day method. However the
amino acids i.e., glutamic acid and glycine were significantly underestimated
as the FM-BOD5 equivalent values were significantly outside the acceptable
range. The effect of incubation time was also investigated in this test. For
the simple sugars (glucose and sucrose), FM-BOD5 equivalent values
slightly increased within the allowable range over 3 h incubation time except
for glucose at 3 h incubation where it gives overestimated value.
Table 2: |
BOD5, FM-BOD5 equivalent values and
percentage degradation of real samples |
 |
For glutamic acid, a significant increase in limiting current over 3 h incubation
time was observed but it is still underestimated compared to the actual BOD5
value. Glycine has shown only a slight increase and underestimated.
Evidently, microbial consortium used in this study does not have the capacity to degrade a broad range of substrates. It is more suitable for simple sugars i.e., glucose and sucrose as substrate. Amino acids i.e., glutamic acid and glycine are found not suitable.
Application of FM-BOD assay to industrial wastewater: The FM-BOD assay
employing the microbial consortium was applied to real industrial wastewater
and river water samples (Table 2) with BOD values of 1-120000
mg BOD5/L. It has been reported that the linearity of the FM-BOD
method is approximately 200 mg L-1 for the standard GGA of solution
(Morris et al., 2001). Hence, the industrial
wastewater sample used in this study was diluted accordingly prior to FM-BOD
measurement.
Table 2 presents the conventional BOD5 values, FM-BOD5 equivalent values and percentage degradation for the microbial consortium in the diluted waste waters. FM-BOD5 equivalent values were obtained after 1 h of incubation. The percentage degradation of food wastewater samples such as pineapple and cafeteria 1 and 2 in 1 h incubation FM-BOD assay were 93, 49 and 71%, respectively which compared favorably with the GGA standard solution BOD5 assay (~60% GGA degraded in 5 days). On some occasions, food waste such as coffee has shown significant underestimated BOD5 equivalent value in 1 h incubation FM-BOD assay. Other industrial wastewaters such as palm oil mill effluent, POME (raw and treated), textile and petrochemical were also applied in the FM-BOD assay. The percentage degradation were all underestimated which are only 5-10% (data not shown). These types of wastewater were hardly assimilable and the microbial consortium used was not able to degrade the compounds in such a short time. Similar to previous discussion microbial consortium is more suitable for degrading food wastewater samples with simple compounds such as glucose and sucrose. Therefore, this FM-BOD assay with the microbial consortium is found to be more suitable for wastewaters that are related to food with easily degradable carbon source.
Some river waters from different places were taken and applied on the FM-BOD
assay. BOD5 values of river waters are usually <10 mg L-1
and contain less biodegradable organic compounds such as humic acid, lignin,
tannic acid, gum arabic and surfactants (Chee et al.,
1999). River water with its BOD5 values <10 mg L-1
is thus not suitable as it is limited by the sensitivity of this FM-BOD assay.
CONCLUSION
The present study has shown that a locally developed microbial consortium can
be used as biocatalyst in the FM-BOD assay for rapid BOD determination. A commercial
consortium was unable to utilize ferricyanide as the final electron acceptor
compared with locally developed microbial consortium in FM-BOD assay. All experiments
in this study were conducted by using microbial consortium with each bacterium
mixed at the same optical density (OD600) of 5 and incubated with
60 mM of ferricyanide for 1 h with GGA as the standard solution. FM-BOD assay
was applied on synthetic organic solutions as well as real wastewater samples.
In conclusion, BOD of food samples with easily degradable carbon source can
be determined within 1 h using FM-BOD assay compared to 5 days standard BOD5
assay.
ACKNOWLEDGMENT
The authors would like to acknowledge the Ministry of Science, Technology and Innovation of Malaysia (Vote 79156) for financial support of this study.