Use of Modified Nano Sorbent Materials for Extraction and Determination of Trace Amounts of Copper Ions in Food and Natural Water Samples
|
|
A. Mirabi
and
S.N. Hosseini
|
| |
ABSTRACT
|
|
In this research, the first study on the high performance of Nanoparticled γ-alumina coated with sodium dodecyl sulfate and 4-2-pyridylazo-resorcinol (SDS-PAR) as a new sorbent solid phase extraction has been reported. Sorbent with adjusted nanometer-sized alumina was used to preconcentrate and separate Cu in plant belonging to the legume family and natural water samples. A microsample introduction system was employed for the nebulization micro-volume of diluted solution by FAAS (flame atomic absorption spectrometry) method. Various influencing parameters on the separation and preconcentration of trace metal, such as pH, amount of sorbent, and volume of the chelating agent, were studied. The calibration graph was linear in the range of 1-800 μg L-1 with detection limit of 0.21 μg L-1. RSD for ten reproduce measurements was 1.8%. The proposed method has been applied to the decision of Cu ions at trace levels in substantial samples such as cucumber, eggplant, mint, tomato, potato, parsley, spinach, apple, mangosteen, kiwi, banana, The apple core, macaroni, pea, wheat flour, red beans, lentil, barley, tap water, river water, sea water. The proposed analytical method was successfully applied to the determination of Cu in mentioned samples. |
|
| |
|
|
| |
| Received:
November 05, 2011; Accepted: February 17, 2012;
Published: May 25, 2012 |
|
|
INTRODUCTION
The determination of trace metal elements, mainly intermediate metals has received
increasing attention in pollution studies. Copper is an essential trace element
at lower concentrations in biological systems. In view of plenty of it in universe,
investigation of copper in various samples has been studied, so developing sensitive
and selective methods to determine copper are required. In general, a daily
copper intake of 1.5-2 mg is essential and copper at nearly 40 ng mL-1
is required for biological processes. But, in higher concentration, it is toxic
and severe oral intoxication will affect mainly the blood and kidneys. In these
levels, copper can bind to the cell membrane and hinder the transport process
through the cell wall (Ansari et al., 2004; Ghiasvand
et al., 2004), so they affect on the taste of waters and corrosion.
The copper content at trace levels in water and food must be controlled thus
the European Commissioner, United States and Canada have fixed amount of Cu
in drinking water, respectively 2 g mL-1, 1.3 and 1.0 μg mL-1
(Rumori and Cerda, 2003; Szigeti
et al., 2005). FAAS selectivity is good and it has high sensitivity,
speed, satisfactory reproducibility, wide dynamic concentration range and low
cost (Soylak, 2004; Awode et
al., 2008; Pantuzzo et al., 2009; Khaniki
et al., 2007; Ezer, 2009). Anyway, determination
of metals at sub-mg L-1 level by FAAS is not directly possible (Soylak
et al., 2005; Calero et al., 2009;
Rofouei et al., 2009; Iwegbue,
2012; Ojeda and Rojas, 2009). Also another problem
in these determinations is interference of the matrix constituents of the analyzed
samples (Soylak and Elci, 2000; Khosravan
and Saljooghi, 2009; Tharanitharan and Srinivasan, 2009;
Bakircioglu et al., 2009). Various separation-preconcentration
methods encompassing solvent extraction, cloud point extraction, ion-exchange,
electroanalytical techniques, solid phase extraction, and chelating sorbents,
etc. (Soylak and Tuzen, 2006; Burham
et al., 2009; Quiros, 2009; Soylak
and Dogan, 1996; Rezaei et al., 2009; Shamspur
and Mostafavi, 2009; Khammas, 2009; Rappoport
and Liebman, 2009) have been used to overcome these problems. Among these
sorbents, Nanoscale materials have gained great attention because they have
special properties, because nanometer-sized alumina has high surface area, high
sorption capacity and high chemical activity and it has ability for the separation
and preconcentration of trace elements in biological environmental samples.
Recently, the using of the surfactant-coated mineral oxide for SPE has demonstrated
that for these new sorbent materials to be a favorable tool for the extraction/preconcentration
of organic compounds in a wide polarity range. In this study, FAAS has been
applied for the preconcentration and determination of Cu in real samples. Nanometer-sized
alumina coated with SDS-PAR has been used as a new and effective sorbent. High
efficiency has been shown.
MATERIALS AND METHODS Materials: Usually in chemistry, mainly analytical chemistry, analytical-reagent grade is used. The model compound studied were Cu(II) nitrate(Merck), nano-γ-alumina (Sigma-Aldrich), nitric acid (Merck), SDS (Merck) and 4-(2-Pyridylazo) resorcinol monosodium salt monohydrate (PAR) (Merck). They were used without further purification. All solutions were prepared in distilled water. Instrumentation: Atomic absorption analyses were performed on an M series model (Model: M5) equipped with a 50 mm burner head, deuterium background correction and using an air-acetylene flame and resonance line of 324.8 nm for copper. The buffer solutions have been applied to adjust the pH of the solutions. An expanded scale pH meter Metrohm (model 744) equipped with glass electrode for measurements. A centrifuge (Kokusan Ensinkico, model H-11N) is used to accelerate the phase separation process.
Procedure of modified nano-γ-alumina: Ten milliliter solution with
pH of 2 contains 0.5 g of activated nano-γ-alumina, 0.1 mg SDS and 0.2
mL PAR. Then it was shaked for 10 min (Shemirani and Akhavi,
2001).
General procedure: The sample solution of Cu (II) in the range of 1-800 μg L-1 was added to modified nano-γ-alumina and the adjusted pH value was 4 for using acetic acid buffer solution. The centrifuge was used in the process of solid phase extraction at 3800 rpm for 10 min. Then, for eluting of the metal ions retained on the sorbent, were used 0.2 mL of HNO3 1 mol L-1. The determination of analytes was used FAAS in the extraction.
Preparation of real samples: Cucumber (2.0 g), eggplant (2.0 g), mint
(2.0 g), tomato (2.0 g), potato (2.0 g), parsley (2.0 g), spinach (2.0 g), apple
(3.0 g), mangosteen (2.0 g), kiwi (2.0 g), banana (2.0 g), pea (2.0 g), The
apple core (2.0 g), macaroni (2.0 g) wheat flour (2.0 g), red beans (2.0 g),
lentil (2.0 g) and barley (2.0 g) samples were digested with 10 mL of HNO3
(65%) and 3 mL of H2O2 (30%) . Then after this action,
the volume of the digested samples was made up to 25 mL with distilled water.
RESULTS AND DISCUSSION Investigation of all the parameters is necessary because they can influence the extraction performance. In this method, these parameters are the kind and amount of eluting solution and anti-sticking pageant, pH, Ligand amount, centrifuge conditions that were investigated and optimized in order to achieve a high recovery and enrichment factor. In all optimization steps concentration of copper was 0.5 mg L-1.
Effect of pH: The pH of the solutions is very important in the SPE procedure.
The sorption of SDS on nano-sized alumina is highly dependent on the pH of solution.
Positively charged nano-γ-alumina surfaces effectively sorbed negatively
charged SDS at lower pH values. Therefore, maintenance of organic compounds
on SDS-coated nano-γ-alumina occurs. The SDS is absorbed on the positively
charged nano-γ-alumina surfaces via formation of self-aggregates (Ghaedi
et al., 2009; Li et al., 2008) over
a wide pH range (1-6), whereas very little amount of SDS could be retained on
inert surface of nano-γ-alumina. Maximum sorption of SDS on nano-γ-alumina
was achieved at pH 1-2.9 by shaking the solution containing SDS and nano-γ-alumina
for 10 min. When solution was acidified, SDS would form hemimicelles on nano-γ-alumina
by strong sorption. The micelles could trap PAR molecules homogeneously (Fig.
1). (The low toxicological properties, commercial availability and cost
are very important factors in SDS) (Shemirani and Akhavi,
2001).
The pH of the aqueous solution is an important parameter for retaining of metal ions on a solid phase materials as metal ions-chelates. The sorption behavior of Cu by modified nano-γ-alumina in water samples was investigated in the pH range of 3-9. According to the results, a quantitative recovery of Cu (>95%) was found in the pH 4. So pH 4 was used in this working (Fig. 2). Effect of the amount of sorbent: Regard to the experimental results (Fig. 3), copper ion was completely adsorbed on the sorbent in the 400 mg of the modified nano-γ-alumina. Therefore, 400 mg was selected for all of the subsequent experiments.
|
| Fig. 1: |
A suggested binding mode of SDS-PAR to nano-alumina surface |
|
| Fig. 2: |
Effect of pH on Cu sorption |
|
| Fig. 3: |
Effect of the amount of modified nano-γ-alumina on Cu
sorption |
Effect of eluting solution condition: In this experiment, a series of experiments were designed and performed to obtain a reasonable eluent to elute completely Cu2+ ions after their enrichment by chelation with modified nano-γ-alumina. Various acids with various concentrations (0.1-1 mol L-1) were to obtain the optimal eluent for further experiments. The results showed that HNO3 0.1 mol L-1 accomplish the quantitative elution of target analyte, Cu2+ ions chelated with from the PAR in solid phase extraction (Table 1). Effect of volume of ligand: The effect of PAR volumes in the range of 0.05-4 mL was evaluated on the recoveries of analytes. Figure 4 shows that the extraction recovery increases up to 1.0 mL of ligand and then remains constant. So, 1.0 mL of ligand 10-3 M was chosen as optimum. Effect of centrifuge conditions: The effect of centrifugation rate on the absorbance was studied in the range of 1000-4000 rpm. It was found when rpm was over 3000, solid phase completely settled, so the optimizing of rate was selected 3800 rpm. The effects of centrifugation time were investigated on the settles of solid phase in the range of 5-15 min at 3800 rpm. Over 7 min, transfer of solid phase to the bottom of centrifuge tube was complete. So, the optimum centrifugation time was chosen as 10 min.
|
| Fig. 4: |
Effect of volume of ligand on the sorption of Cu |
| Table 1: |
Effect of type and concentration of eluent on the adsorption
of Cu |
 |
| Table 2: |
Influences of some foreign ions on the recoveries of analytes |
 |
Effect of other ions: The effect of over 13 ions on the determination
of Cu (II) was studied with flame atomic absorption spectrometric. Different
amounts of metal ions were added to a solution containing fixed amounts Cu of
ions and the present procedure was followed. The results for this study are
given in Table 2. The tolerance limit was defined as the ion
concentration causing a relative error smaller than ±5% related to the
coprecipitation and determination of the examined elements.
| Table 3: |
Analytical characteristics of proposed method |
 |
| Table 4: |
Determination of copper in real and spiked water samples |
 |
| aFrom drinking water system of Qaemshahr, Iran,
bTelar river, Qaemshahr, Iran, cCaspian sea water,
Iran, dRSD of three replicate experiments |
Most of the cations and anions examined did not show interferences at the concentration
of 100 times higher than that of Cu concentration, the results are desired in
view of applications to real samples including various salts at high concentration
levels.
Figures of merit: Table 3 shows the analytical features of the optimized method. It contains limit of detection, reproducibility and enhancement factor. The limit of detection (LOD = 0.21 μg L-1) was calculated as 3Sb/m (Sb: standard deviation of the blank signals. m: slope of calibration curve after preconcentration). A reasonable correlation coefficient (r2 = 0.999) was gotten and only small deviations between sequential determinations (RSD = 1.8%) were found. The calibration curve was investigated up to 2000 μg L-1 which was linear. Enhancement factor (EF = 50) was obtained from the slope ratio of calibration curve after and before preconcentration. The calibration equation curve after the preconcentration procedure was given in Eq. 1. Analysis of real sample: This method was applied to determine of Cu in water and herbal samples. In order to test the applicability and reliability of the proposed method, tap water, river water, sea water, cucumber, eggplant, mint, tomato, potato, parsley, spinach, apple, mangosteen, kiwi, banana, pea, The apple core, macaroni, wheat flour, red beans, lentil and barley samples were analyzed. For this purpose, 10 mL of each of the samples were preconcentrated with modified nano-γ-alumina, in accordance with the proposed method. The results are shown in Table 4 and 5. In all cases the spike recoveries confirmed the reliability of the proposed method.
Comparison other SPE studies: The data have been compared with those
of recent reported methods on preconcentration of copper (Table
6). Some parameters obtained are comparable to those presented by other
methods. The present work has relatively low LOD and RSD when compared to other
sorbent, activated carbon, silica gel, resine.
| Table 5: |
Determination of copper in herbal samples |
 |
| aRSD of three replicate experiments |
| Table 6: |
Comparison of the proposed method with other methods reported |
 |
Other parameter, PF (50) is relatively high enough when compared to some of
the others methods. In addition, the cited methods are more complicated and
use more reagents than this method, namely application of this method is simpler
and takes less time.
CONCLUSION For determination of trace amounts of copper ions in low concentrations, the ligand loaded on the modified nano-γ-alumina was used. These results showed that this method has high sensitivity. The proposed solid phase extraction method has important advantages such as: it is superior for having lower RSD and lower detection limits when compared to other methods (Table 6). It is simple, rapid with low cost. Therefore, this method of the determination of copper by FAAS provides a simple, easy and accurate for the preconcentration and separation of trace amounts. The method has been successfully applied to the environmental samples. ACKNOWLEDGMENT The author thanks the research council at the Qaemshahr Branch of Islamic Azad University for financial support.
|
|
REFERENCES |
Ansari, T.M., N. Ikram, M. Najam-ul-Haq, I. Fayyaz, Q. Fayyaz, I. Ghafoor and N. Khalid, 2004. Essential trace metal (Zinc, Manganese, Copper and Iron) levels in plants of medicinal importance. J. Biol. Sci., 4: 95-99. CrossRef | Direct Link |
Awode, U.A., A. Uzairu, M.L. Balarabe, G.F.S. Harrisson and O.J. Okunola, 2008. Assessment of peppers and soils for some heavy metals from irrigated farmlands on the bank of river challawa, Northern Nigeria. Pak. J. Nutr., 7: 244-248. CrossRef | Direct Link |
Bakircioglu, D., M. Kuscu and Y. Bakircioglu, 2009. Online solid phase preconcentration using EGDMA-MAA copolymer for lead determination by FAAS. Toxicol. Environ. Chem., 91: 1205-1215. CrossRef |
Burham, N., S.M. Abdel-Azeem and M.F. El-Shahat, 2009. Removal of Pb and Cd from aqueous media and fish liver using novel polyurethane foam functionalized with pyrazolone as a new metal ion collector. Cent. Eur. J. Chem., 7: 576-585. CrossRef |
Calero, M., F. Hernainz, G. Blazquez, M.A. Martin-Lara and G. Tenorio, 2009. Biosorption kinetics of Cd (II), Cr (III) and Pb(II) in aqueous solutions by olive stone. Braz. J. Chem. Eng., 26: 265-273. Direct Link |
Durukan, I., C.A. Aahin and S. Bektas, 2011. Determination of copper tracts in water samples by flow injection-flame atomic absorotion spectrometry using a novel solidified floatingorganic drop microextraction method. Microchem. J., 98: 215-219.
Ezer, M., 2009. Heavy metal content of roadside soil in Kahramanmaras, Turkey. Fresenius Environ. Bull., 18: 704-708. Direct Link |
Ferreira, S.L.C., M.A. Bezerra, W.N.L. Dos Santos and B.B. Neto, 2003. Application of doehlert designs for optimization of an on-line preconcentration system for copper determination by flame atomic absorption spectrometry. Talanta, 61: 295-303.
Ghaedi, M., H. Tavallali, A. Shokrollahi, M. Zahedi, M. Montazerozohori and M. Soylak, 2009. Flame atomic absorption spectrometric determination of zinc, nickel, iron and lead in different matrixesafter solid phase extraction on sodium dodecyl sulfate (SDS) coated alumina as their bis(2-hydroxyacetophenone)-1,3-propanediimine chelates. J. Hazard. Mater., 166: 1441-1448. Direct Link |
Ghiasvand, A.R., R. Ghaderi and A. Kakanejadifard, 2004. Selective preconcentration of ultra trace copper(II) using octadecyl silica membrane disks modified by a recently synthesized glyoxime derivative. Talanta, 62: 287-292. PubMed |
Iwegbue, C.M.A., 2012. Metal contents in some brands of biscuits consumed in southern Nigeria. Am. J. Food Technol., 7: 160-167. CrossRef |
Khammas, Z.A.A., 2009. Recent trends for separation and preconcentration in metal ions and organic compounds analysis after Clould-point methodology: Developments and analytical applications: A review. Eurasian J. Anal. Chem., 4: 1-35. Direct Link |
Khaniki, G.R.J., M.H. Dehghani, A.H. Mahvi and S. Nazmara, 2007. Determination of trace metal contaminants in edible salts in Tehran (Iran) by atomic absorption spectrophotometry. J. Boil. Sci., 7: 811-814. CrossRef | Direct Link |
Khosravan, M. and A.S. Saljooghi, 2009. Synthesis, characterization of silica gel phases chemically immobilized with-2-aminothiophenol and the use of it for separation and preconcentration of trace amounts of manganese ions. Eur. J. Scient. Res., 30: 58-66. Direct Link |
Li, J.D., Y.L. Shi, Y.Q. Cai, S.F. Mou and G.B. Jiang, 2008. Adsorption of di-ethyl-phthalate from aqueous solutions with surfactant-coated nano/microsized alumina. Chem. Eng. J., 140: 214-220. Direct Link |
Ojeda, C.B. and F.S. Rojas, 2009. Separation and preconcentration by a cloud point extraction procedure for determination of metals: An overview. Anal. Bioanal. Chem., 394: 759-782. CrossRef |
Pantuzzo, F.L., J.C.J. Silva and V.S.T. Ciminelli, 2009. A fast and accurate microwave-assisted digestion method for arsenic determination in complex mining residues by flame atomic absorption spectrometry. J. Hazard. Mater., 168: 1636-1638. PubMed |
Quiros, M.B., 2009. Comparison of analytical methodologies for aluminium determination in hemodialysis fluids. Portugaliae Electrochim. Acta, 27: 69-76. Direct Link |
Rappoport, Z. and J.F. Liebman, 2009. The Chemistry of Hydroxylamines, Oximes and Hydroxamic Acids, Volume 1. John Wiley and Sons, New York, ISBN: 9780470512616, Pages: 1038.
Rezaei, B., E. Sadeghi and S. Meghdadi, 2009. Nano-level determination of copper with atomic absorption spectrometry after pre-concentration on N, N-(4-methyl-1,2- phenylene) diquinoline-2-carboxamide-naphthalene. J. Hazard. Mater., 168: 787-792. PubMed |
Rofouei, M.K., M. Payehghadr, M. Shamsipur and A. Ahmadalinezhad, 2009. Solid phase extraction of ultra traces silver (I) using octadecyl silica membrane disks modified by 1, 3-bis(2-cyanobenzene) triazene (CBT) ligand prior to determination by flame atomic absorption. J. Hazard. Mater., 168: 1184-1187. Direct Link |
Rumori, P. and V. Cerda, 2003. Reversed flow injection and sandwich sequential injection methods for the spectrophotometric determination of copper(II) with cuprizone. Anal. Chim. Acta, 486: 227-235. Direct Link |
Shamspur, T. and A. Mostafavi, 2009. Application of modified multiwalled carbon nanotubes as a sorbent for simultaneous separation and preconcentration trace amounts of Au(III) and Mn(II). J. Hazard. Mater., 168: 1543-1553. Direct Link |
Shemirani, F. and B.T.S. Akhavi, 2001. Preconcentration and determination of trace cadmium using 1-(2-pyridylazo)-2-naphthol (PAN) immobilized on surfactant-coated alumina. Anal. Lett., 34: 2179-2188. CrossRef |
Soylak, M. and L. Elci, 2000. Solid phase extraction of trace metal ions in drinking water samples from Kayseri-Turkey. J. Trace Microprobe Tech., 18: 397-403.
Soylak, M. and M. Dogan, 1996. Separation and enrichment of zinc, iron, copper, cadmium, cobalt and nickel from urine with amberlite XAD-16 resin. Trace Elem. Electroly, 13: 130-132.
Soylak, M. and M. Tuzen, 2006. Diaion SP-850 resin as a new solid phase extractor for preconcentration-separation of trace metal ions in environmental samples. J. Hazard. Mater., 137: 1496-1501. Direct Link |
Soylak, M., 2004. Solid phase extraction of Cu(II), Pb(II), Fe(III), Co(II) and Cr(III) on chelex 100 column prior to their flame atomic absorption spectrometric determinations. Anal. Lett., 37: 1203-1217.
Soylak, M., S. Saracoglu, M. Tuzen and D. Mendil, 2005. Determination of trace metals in mushroom samples from Kayseri, Turkey. Food Chem., 92: 649-652. CrossRef | Direct Link |
Szigeti, Z., I. Bitter, K. Toth, C. Latkoczy, D.J. Fliegel, D. Gunther and E. Pretsch, 2005. A novel polymeric membrana electrode for the potentiometric anylisis of Cu2+ in drinking water. Anal. Chim. Acta, 532: 129-136. Direct Link |
Tabrizi, B.A., 2007. Development of cloud point extraction-spectrofluorimetric method for trance copper (II) determination in water samples and parenteral solutions. J. Hazard. Mater., 139: 260-264.
Tharanitharan, V. and K. Srinivasan, 2009. Removal of Ni(II) from water and wastewater using modified Duolite XAD-761 resin. Indian J. Chem. Technol., 16: 245-253. Direct Link |
Zeeb, M., M.R. Ganjali, P. Norouzi and M.R. Kalaee, 2011. Sepration and preconcentration system based on microextraction with ionic liquid determination of copper in water and food samples by stopped-flow injection spectrofluorimetry. Food Chem. Toxicol., 49: 1086-1091. Direct Link |
|
|
|
 |